  1. @TABLE HEADING = SAMPLE TROUBLE SHOOTING TABLE<$ITable Heading Tag><$ITags;Table 
  2. Heading>
  3.  
  4. @T2 C1 HEAD = Problem<$IT2 C1 Head Tag><$ITags;T2 C1 Head>
  5.  
  6. @T2 C2 HEAD = Probable Cause<$IT2 C2 Head Tag><$ITags;T2 C2 Head>
  7.  
  8. @T2 C3 HEAD = Possible Solutions<$ITags;T2 C3 Head>
  9.  
  10. @T2 C1 COPY = Vacuum pump does not lower pressure to ultimate system 
  11. pressure.<$IT2 C1 Copy Tag><$ITags;T2 C1 Copy>
  12.  
  13. @T2 C2 COPY = Oil Contamination<$IT2 C2 Copy Tag><$ITags;T2 C2 Copy>
  14.  
  15. @T2 C3 COPY = Blank-off pump at intake, then inspect pump ultimate 
  16. pressure to determine if oil needs changing.<$IT2 C3 Head Tag><$IT2 
  17. C3 Copy Tag><$ITags;T2 C3 Copy>
  18.  
  19. @T2 C1 COPY = 
  20.  
  21. @T2 C2 COPY = Out-Gassing
  22.  
  23. @T2 C3 COPY = Unusual vapors or materials not normally found wihin 
  24. interia of chamber may be affecting pump performance. These could 
  25. cause the Pirani gauge to register the incorrect pressure even though 
  26. no vacuum leaks exist. Increase the pump down time if necessary to 
  27. compensate for the error.
  28.  
  29. @T2 C1 COPY = 
  30.  
  31. @T2 C2 COPY = Virtual Leak
  32.  
  33. @T2 C3 COPY = Argon gas may be seaping into the interior of the growth 
  34. chamber through faulty valve seal. Alternately, solvents or air entrapments 
  35. may be effecting the results of the <169>Leak Back Test<170>.
  36.  
  37. @T2 C1 COPY = Water Flow Lamp remains on when coolant supply is turned 
  38. off.
  39.  
  40. @T2 C2 COPY = Circuit needs calibration
  41.  
  42. @T2 C3 COPY = Perform the procedure in Chapter 8, <169>Flow Switch 
  43. Adjustment<170> to check the circuit for proper operation. 
  44.  
  45. @T2 C1 COPY = Water Flow Lamp remains off when coolant supply is turned 
  46. on.
  47.  
  48. @T2 C2 COPY = Faulty V1 valve or flow switch paddle
  49.  
  50. @T2 C3 COPY = Perform the procedures in Chapter 8, <169>Flow Switch 
  51. Adjustment<170> to check the circuit for proper operation. 