  1.        Document 0679
  2.  DOCN  M9620679
  3.  TI    Assessment of aerosol containment on the ELITE flow cytometer.
  4.  DT    9602
  5.  AU    Ferbas J; Chadwick KR; Logar A; Patterson AE; Gilpin RW; Margolick JB;
  6.        Department of Infectious Diseases and Microbiology, Graduate; School of
  7.        Public Health, University of Pittsburgh, Pennsylvania,; USA.
  8.  SO    Cytometry. 1995 Mar 15;22(1):45-7. Unique Identifier : AIDSLINE
  9.        MED/96090282
  10.  AB    Biohazardous aerosols generated during cell sorting have been of
  11.        increased concern recently because of interest in sorting specimens
  12.        containing human immunodeficiency virus type 1 (HIV-1). Current flow
  13.        cytometers have features designed to contain such aerosols within the
  14.        sorting chamber, but the efficacy of these features has not been
  15.        established. Therefore, we tested aerosol containment by two ELITE flow
  16.        cytometers (Coulter Cytometry, Inc., Hialeah, FL) during sorting of
  17.        specimens containing high titers of bacteriophage. Agar plates confluent
  18.        with susceptible Escherichia coli were used to detect infectious units
  19.        released from the sorting chamber. Under recommended operating
  20.        conditions very few infectious units were released from the sorting
  21.        chambers. Release increased when the center stream was not optimally
  22.        collected in a vacuum-exhausted tube or the chamber door was not
  23.        completely closed. Failure of the negative pressure and high efficiency
  24.        particle air (HEPA) filtration features had less of an effect. The data
  25.        indicate that these standard safety features provide a rational
  26.        expectation of safety for the flow cytometry operator.
  27.  DE    Aerosols  Cell Separation/INSTRUMENTATION  *Containment of Biohazards
  28.        Evaluation Studies  Flow Cytometry/*INSTRUMENTATION  Human
  29.        HIV-1/ISOLATION & PURIF  Support, U.S. Gov't, P.H.S.  JOURNAL ARTICLE
  30.  
  31.        SOURCE: National Library of Medicine.  NOTICE: This material may be
  32.        protected by Copyright Law (Title 17, U.S.Code).
  33.  
  34.