  1. @TABLE HEADING = SAMPLE TROUBLE SHOOTING TABLE<$ITable Heading Tag><$ITags;Table 
  2. Heading>
  3.  
  4. @T2 C1 HEAD = Problem<$IT2 C1 Head Tag><$ITags;T2 C1 Head>
  5.  
  6. @T2 C2 HEAD = Probable Cause<$IT2 C2 Head Tag><$ITags;T2 C2 Head>
  7.  
  8. @T2 C3 HEAD = Possible Solutions<$ITags;T2 C3 Head>
  9.  
  10. @T2 C1 COPY = Vacuum pump does not lower pressure to ultimate system 
  11. pressure.<$IT2 C1 Copy Tag><$ITags;T2 C1 Copy>
  12.  
  13. @T2 C2 COPY = Oil Contamination<$IT2 C2 Copy Tag><$ITags;T2 C2 Copy>
  14.  
  15. @T2 C3 COPY = Blank-off pump at intake, then inspect pump ultimate 
  16. pressure to determine if oil needs changing.<$IT2 C3 Head Tag><$IT2 
  17. C3 Copy Tag><$ITags;T2 C3 Copy>
  18.  
  19. @T2 C1 COPY = 
  20.  
  21. @T2 C2 COPY = Clogged oxide gettering trap or filter.
  22.  
  23. @T2 C3 COPY = Open gettering trap and clean filter, then remove all 
  24. traces of ioxide powder from interior of cannister.
  25.  
  26. @T2 C1 COPY = 
  27.  
  28. @T2 C2 COPY = Out-Gassing
  29.  
  30. @T2 C3 COPY = Unusual vapors or materials not normally found wihin 
  31. interia of chamber may be affecting pump performance. These could 
  32. cause the Pirani gauge to register the incorrect pressure even though 
  33. no vacuum leaks exist. Increase the pump down time if necessary to 
  34. compensate for the error.
  35.  
  36. @T2 C1 COPY = 
  37.  
  38. @T2 C2 COPY = Virtual Leak
  39.  
  40. @T2 C3 COPY = Argon gas may be seaping into the interior of the growth 
  41. chamber through faulty valve seal. Alternately, solvents or air entrapments 
  42. may be effecting the results of the <169>Leak Back Test<170>.
  43.  
  44. @T2 C1 COPY = Water Flow Lamp remains on when coolant supply is turned 
  45. off.
  46.  
  47. @T2 C2 COPY = Circuit needs calibration
  48.  
  49. @T2 C3 COPY = Perform the procedure in Chapter 8, <169>Flow Switch 
  50. Adjustment<170> to check the circuit for proper operation. 
  51.  
  52. @T2 C1 COPY = Water Flow Lamp remains off when coolant supply is turned 
  53. on.
  54.  
  55. @T2 C2 COPY = Faulty V1 valve or flow switch paddle
  56.  
  57. @T2 C3 COPY = Perform the procedures in Chapter 8, <169>Flow Switch 
  58. Adjustment<170> to check the circuit for proper operation. 
  59.  